OOM202 氧气传感器
产品品牌:安维特ENVITEC
产品型号:OOM202
所属分类:
PM 2.5传感器 二氧化碳传感器 甲醛传感器 氧气传感器 硫化氢 酒精 CITY\DDS\FIGARO\NEMOTO\solidsense\SST\honeywe\Alphasense\Membrapor\Baseline-mocon\Southland\IT\SGX气体传感器
品牌:安维特ENVITEC 型号:OOM202
可适用于PB760/PB840呼吸机
AVEA呼吸机
天马呼吸机
哈美顿呼吸机
阿拉丁呼吸机
等
Measurement Range: 0-100 % oxygen
Output in ambient air: 13 to 16mV
Electrical Interface: 3pin (Molex 22-11-1031)
Accuracy and Repeatability: < 1 % vol. O2 when calibrated at 100 %
Oxygen
Linearity error: < 3 % relative
Response time: < 12 sec. to 90 % of final value
Zero Offset Voltage: < 200 μV in 100 % nitrogen
applied for 5 min
Cross Interference: < 0.5 % vol. O2 response to:
10 % CO2 balance N2
80% N2O balance N2
7.5% Halothane balance N2
7.5 % Isoflurane balance N2
7.5 % Enflurane balance N2
9% Sevoflurane balance N2
20% Desflurane balance N2
Influence of Humidity: - 0.03 % rel. per % RH at 25°C
Influence of Pressure: proportional to change in oxygen partial
pressure
Influence of Mechanical Shock: < 1% relative after a fall from 1m
Operating Temperature: 0 to 50°C
Temperature Compensation: built-in NTC compensation
Effect of Temperature
Compensation (steady state):
between +25°C and +40°C: 3 % relative error
between 0 °C and +50 °C: 8 % relative error
Operating Humidity: 0-99 % RH non-condensing
Long Term Output Drift: < 1 % vol oxygen per month
typically < - 15 % relative over lifetime
Storage Temperature: -20 to +50 °C
Recommended Storage: +5 to +15 °C
Recommended Load: ³ 10 kOhms
Warm-Up Time: < 30 minutes, after replacement of sensor
Nominal Sensor Lifetime: ³ 1.000 000 % vol oxygen hours
Weight: approximately 28 grams
Part No.: 01-00-0047